TB 9-6625-2241-24
13. Power Supply
NOTE
Do not perform power supply checks if all parameters are
within tolerance.
a. Performance Check. Connect multimeter to TI TEST POINT +65V (fig. 2) and
chassis ground.  If digital multimeter does not indicate between +64.67 and +65.33 V
perform b below.
b. Adjustments. Adjust R1736 +65V ADJ (fig. 2) until multimeter indicates +65V (R).
14. Final Procedure
a. Deenergize and disconnect all equipment and reinstall protective cover on TI.
b. Annotate and affix DA Label/Form in accordance with TB 750-25.
17/(18 Blank)