SECTION III. CALIBRATION PROCESS
Figure 2. UnIeveled output power gain - equipment setup.

Calibration Procedure For Microwave Amplifier Model 8349B Page Navigation
  1    2    3  4  5    6    7    8    9  
img
. . . . .
TB 9-6625-2249-24
Figure 1. Receiver system reference equipment setup.
g. Configure measuring receiver to measure frequency at 2 GHz.
h. Configure measuring receiver to measure power and establish a 0.00 dB reference.
NOTE
Verify the proper cal factors are loaded for the power sensor
module being utilized.
8. Unlevel Output Power Gain
a. Performance Check
NOTE
Ensure measuring receiver indicates between 0.00 and 0.01 dB
in power measurement mode before proceeding to (1) below.
(1) Connect equipment as shown in figure 2.
(2) Adjust synthesized signal generator level controls for a -5 dBm indication on
power meter.
(3) Configure measuring receiver to measure power. Indication will be equal to or
greater than 15 dB.
(4) Adjust synthesized signal generator frequency controls for a display of first test
frequency listed in table 3.
5