• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
Figure 1. Receiver system reference equipment setup.
Figure 3. Leveled out output power gain and flatness - equipment setup.

Calibration Procedure For Microwave Amplifier Model 8349B Page Navigation
  1    2    3    4  5  6    7    8    9    10  
img
. . .
TB 9-6625-2249-24
Figure 2. UnIeveled output power gain - equipment setup.
Table 3. Unleveled Gain Test Frequencies
Test frequencies (GHz)
3.0
5.6
9.7
11.0
15.0
17.0
18.0
(5) Configure measuring receiver to measure frequency. Indication will be the same
as synthesized signal generator frequency in (4) above.
(6) Configure measuring receiver to measure power. Indication will be equal to or
greater than 15 dB.
(7) Repeat technique of (2) through (6) above for remaining test frequencies listed in
table 3.
(8) Adjust synthesized signal generator frequency controls for a display of 2 GHz.
(9) Adjust synthesized signal generator level output controls for a +5 dBm
indication on power meter.