Table 4. Leveled Output Power and Flatness
Leveled Output Power Gain and flatness

Calibration Procedure For Microwave Amplifier Model 8349B Page Navigation
  3    4    5    6    7  8  9    10    11    12    13  
img
. . . . . . . .
TB 9-6625-2249-24
Figure 4. Compensation zero adjustments - equipment setup.
(2) Adjust synthesized signal generator frequency controls to 11 GHz and adjust
level output controls for -25 dBm indication on power meter. Record synthesized signal
generator level indication.
(3) Connect equipment as shown in figure 1.
(4) Center A4R60 through A4R67 (fig. 5).
Figure 5. Signal conditioning board - adjustments location.
(5) Connect multimeter negative lead to A4TP4 (fig. 5) and positive lead to A4TP5 (fig. 5).
(6) Adjust A4R67 (fig. 5) for a multimeter indication between -0.001 and +0.001 V dc.
(7) Connect multimeter positive lead to A4TP3 (fig. 5).
9