SECTION I. IDENTIFICATION AND DESCRIPTION
Table 2. Minimum Specifications of Equipment Required

Calibration Procedure For Sweep Generator Sg-1206( )/U Page Navigation
  1  2  3    4    5    6    7  
img
. . . . . . . .
TB 9-6625-2259-24
Table 1. Calibration Description - Continued
Test instrument parameters
Performance specifications
Output power
Frequency range: 10 MHz to 2 GHz
Output power: At least + 13 dBm
Frequency range: 2 to 18 GHz
Output power: At least + 10 dBm
Frequency range: 18 to 20 GHz1
Output power: At least +7 dBm
Output power variation
Frequency range: 10 MHz to 2 GHz
Maximum power internal leveled:  1.0 dB
Maximum power external leveled:  0.3 dB
Frequency range: 2 to 20 GHz1
Maximum power internal leveled: 1.1 dB
Maximum power external leveled: 0.4 dB
Flatness
Frequency range: 10 MHz to 20 GHz
Attenuation range: 0 dBm
Flatness:  1.5 dB
Output attenuator
Frequency range: 10 MHz to 2 GHz
Attenuator range: +13 to - 60 dBm
Total maximum error: 2.6 dB
Attenuator range: -60 to - 100 dBm
Total maximum error: 3.1 dB
Frequency range: 2 to 18 GHz
Attenuator range: 10 to - 60 dBm
Total maximum error: 2.6 dB
Frequency range: 18 to 20 Ghz1
Attenuator range: 7 to -60 dBm
Total maximum error:  2.6 dB
1Not
checked above 18 GHz due to TI output connector limitations.
Checked to 17.5 GHz.
2
SECTION II
EQUIPMENT REQUIREMENTS
4. Equipment required. Table 2 identifies the specific equipment to be used in this
calibration procedure.  This equipment is issued with Secondary Transfer Calibration
Standards Sets AN/GSM-287, AN/GSM-705 or Secondary Reference Set NSN 4931-00-621-7878.
Alternate items may be used by the calibrating activity.  The items selected must be
verified to perform satisfactorily prior to use and must bear evidence of current calibration.
The equipment must meet or exceed the minimum use specifications listed in table 2. The
accuracies listed in table 2 provide a four-to-one-one ratio between the standard and TI.
Where the four-to-one ratio cannot be met, the actual accuracy of the equipment selected is
shown in parenthesis.
5. Accessories Required.  The accessories required for this calibration are common
usage accessories, issued as indicated in paragraph 4 above, and are not listed in this
calibration procedure. The following peculiar accessory is also required for this calibration:
Crystal detector, Hewlett-Packard, Model 423AOPT03 (423AOPT03).