TB 9-6625-2259-24
Table 2. Minimum Specifications of Equipment Required
Manufacture and model
Common Name
Minimum use specifications
(part number)
ATTENUATOR (FIXED)
Frequency range: 10 MHz to 18 GHz
Weinschel, Model 9918,
10 dB attenuator (fixed)
9918-10dB, and 9918-30dB
Accuracy: .025 dB/10 dB (determined in
(9918)
Calibration process)
30 dB attenuator (fixed)
Measuring Receiver system
MEASURING RECEIVER
N5530S consisting of:
Frequency accuracy: 0.02%
Spectrum Analyzer
Residual AM frequency range: 10 MHz to
Agilent, Model E4440A
17.5 GHz
(E4440A), Power Meter
AM <0.32%
Agilent, Model E4419B
Residual FM frequency range: 10 MHz to
(E4419B), and Sensor
17.5 GHz
Module Agilent, Model
FM <5.0 kHz
N5532A opt 518 (518)
Attenuation range: -1.5 to + 15 dBm
MICROWAVE
Range: 10 Hz to 20.01 GHz
Anritsu, Model MF2414B
Accuracy: 0.0125%
FREQUENCY COUNTER
(MF2414B)
Range: -0.19 to 18.2 V dc
Hewlett-Packard, Model
3458A (3458A)
Accuracy:
0.25%
POWER METER
Range: -10.0 to +5 dBm
Hewlett-Packard, Model
Frequency range: 10 MHz to 18 GHz
E12-432A (MIS-30525)
Accuracy: 0.7% (with test report)
w/thermistor mount,
Hewlett-Packard, Model
H75-478A (7915907) or
8478B (8478B)
SPECTRUM ANALYZER
Frequency range: 10 MHz to 18 GHz
(AN/USM-677)
10 dB attenuator (fixed)
Able to view signals to -45 dBc
6. Preliminary Instructions
a. The instructions outlined in paragraphs 6 and 7 are preparatory to the calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
c. Unless otherwise specified, verify the results of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure. Additional
maintenance information is contained in TM 11-6625-3231-40.
d. Unless otherwise specified, all controls and control settings refer to the TI.