Frequency Accuracy Test - Continued
Frequency Accuracy Test - Continued

Calibration Procedure For Sweep Generator Sg-1206( )/U Page Navigation
  2    3    4    5    6  7  8    9    10    11    12  
img
. . .
TB 9-6625-2259-24
(13) Adjust A6R12 (fig. 1) for a microwave frequency counter indication between
2100 and 2102 MHz (R).
Figure 1. A4A4A6 component location diagram.
(14) Replace TI top cover and wait 5 minutes before proceeding to (15) below.
(15) Press FREQUENCY RANGE CW F2 key.
(16) Remove TI top cover.
(17) Adjust A6R6 (fig. 1) for a microwave frequency counter indication between 7898
and 7900 MHz (R).
(18) Replace TI top cover and wait 5 minutes before proceeding to (19) below.
(19) Repeat (8) through (18) above until both FREQUENCY RANGE CW F1 and
CW F2 frequencies are within the specified limits.
8