• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
Frequency Accuracy Test - Continued
Harmonic Attenuation Test

Calibration Procedure For Sweep Generator Sg-1206( )/U Page Navigation
  4    5    6    7    8  9  10    11    12    13    14  
img
.
TB 9-6625-2259-24
WARNING
Perform (40) through (58) below exactly as written to prevent
misalignment  on  wrong  side  of  YIG  oscillator's
hysteresis curve.
(40) Press FREQUENCY RANGE CW F1 key and set F1 M3 display to 12.500
GHz by pressing 12.5 with DATA ENTRY keys and then by pressing GHz/dBm/SEC
DATA ENTRY key.
Figure 2. A4A3A7 component location diagram.
(41) Press FREQUENCY RANGE CW F2 key and set F2 M6 display to 20.000
GHz by pressing 20 with DATA ENTRY keys and then by pressing GHz/dBm/SEC DATA
ENTRY key.