Figure 5. Time base accuracy - adjustment locations.
Table 11. Output Flatness

Calibration Procedure For Audio Analyzer Model 1120-S/10 And 1121 Page Navigation
  10    11    12    13    14  15  16    17    18    19  
img
. . . .
TB 9-6625-2271-40
(5) Replace A5 CPU Y1 (fig. 5) cover screw and TI top cover.
14. Output Level
a. Performance Check
(1) Connect TI OUTPUT HIGH to ac measurement standard INPUT 2 HI and LO.
Setup ac measurement standard to measure levels at INPUT 2.
(2) Press keys and enter values using DATA ENTRY keys as listed in (a) through
(g) below:
LCL INIT.
(a)
SOURCE LEVEL.
(b)
3 mV.
(c)
SOURCE FREQ.
(d)
1 kHz.
(e)
SPCL
(f)
75 ENTER
(g)
(3) If ac measurement standard indication is not between 2.97 and 3.03 mV ac,
perform b below.
(4) Repeat technique of (2) (b) through (g) and (3) above using TI SOURCE settings
and ac measurement standard indications listed in table 10.
Table 10.
Level Accuracy
Test instrument
Ac measurement standard
SOURCE settings
indications
LEVEL
FREQ
Min
Max
3
mV
90  kHz
2.955
mV
3.045  mV
3
mV
120  kHz
2.925
mV
3.075  mV
30
mV
1  kHz
29.835
mV
30.165  mV
30
mV
90  kHz
29.685
mV
30.315  mV
30
mV
120  kHz
29.52
mV
30.48
mV
300
mV
1  kHz
298.35
mV
301.65
mV
300
mV
90  kHz
296.85
mV
303.15
mV
300
mV
120  kHz
295.2
mV
304.8
mV
3
V
1  kHz
2.9835  V
3.0165 V
3
V
90  kHz
2.9685  V
3.0315 V
3
V
120  kHz
2.952
V
3.048  V
16
V
1  kHz
15.912
V
16.088  V
16
V
90  kHz
15.832
V
16.168  V
16
V
120  kHz
15.744
V
16.256  V
(5) Press keys and enter values using DATA ENTRY keys as listed in (a) through
(g) below:
LCL INIT.
(a)
SOURCE LEVEL.
(b)
3 mV.
(c)
SOURCE FREQ.
(d)
1 kHz.
(e)