CAL 06 - Vertical Transient Response - Continued

Calibration Procedure For Oscilloscope Types 2465A And 2465 Page Navigation
  33    34    35    36    37  38
img
. . . . .
TB 9-6625-2294-24
(55) Steps 22 and 23 are automatically calibrated after step 21 and display readout
will indicate: VERT CENTER GAIN.
(56) Adjust GAIN (R638) (fig. 1) and VERTICAL CENTERING (R639) (fig. 1) to align
cursors with the dotted 0 percent and 100 percent graticule markings.
(57) Press and release upper TRIGGER COUPLING pushbutton. Display readout
will indicate: DIAGNSTIC. PUSH A/B TRIG TO EXIT.
22. Final Procedure
a. Press A/B TRIG pushbutton to exit, press POWER pushbutton to off, and move
cal/no cal jumper to "no cal" position and replace protective cover.
b. Repeat paragraphs 7 through 13.