TB 9-6625-2295-24
(41) Adjust ' control for best front corner.
(42) Press and release the upper TRIGGER COUPLING pushbutton. Display
readout will indicate ADJ ' (step) 7, 20 mV, 10 ns.
(43) Set oscilloscope calibrator for an EDGE mode, output of 100 mV at 100 kHz.
Rotate oscilloscope calibrator knob below EDIT FIELD pushbutton to adjust amplitude for
5 divisions of display on TI.
(44) Adjust ' control for best front corner.
(45) Press and release upper TRIGGER COUPLING pushbutton. Display readout
will indicate ADJ ' (step) 8, 50 mV, 10 ns.
(46) Set oscilloscope calibrator for an EDGE mode, output of 250 mV at 100 kHz.
Rotate oscilloscope calibrator knob below EDIT FIELD pushbutton to adjust amplitude for
5 divisions of display on TI.
(47) Adjust ' control for a spike on front corner of 4 to 5 percent.
(48) Press and release upper TRIGGER COUPLING pushbutton. Display readout
will indicate ADJ ' (step) 9, l00 mV, l0 ns.
(49) Set oscilloscope calibrator for an EDGE mode, output of 500 mV at 100 kHz.
Rotate oscilloscope calibrator knob below EDIT FIELD pushbutton to adjust amplitude for
5 divisions of display on TI.
(50) Adjust ' control for a front corner overshoot of 6 to 7 percent. This is necessary
(51) Press and release upper TRIGGER COUPLING pushbutton. Display readout
will indicate ADJ ' (step) 10, 200 mV, 10 ns.
(52) Set oscilloscope calibrator for an EDGE mode, output of 1 V at 100 kHz. Rotate
oscilloscope calibrator knob below EDIT FIELD pushbutton to adjust amplitude for 5
divisions of display on TI.
(53) Adjust ' control for best corner.
(54) Press and release upper TRIGGER COUPLING pushbutton. Display readout
will indicate ADJ ' (step) 11, 500 mV, 10 ns.
(55) Set oscilloscope calibrator for an EDGE mode, output of 2.5 V at 100 kHz.
Rotate oscilloscope calibrator knob below EDIT FIELD pushbutton to adjust amplitude for
5 divisions of display on TI.
(56) Adjust ' control for best corner.
(57) Press and release upper TRIGGER COUPLING pushbutton. Display readout
will indicate ADJ ' (step) 12, 1 V, 10 ns.
(58) Set oscilloscope calibrator for an EDGE mode, output of 2.5 V at 100 kHz.
Rotate oscilloscope calibrator knob below EDIT FIELD pushbutton to adjust amplitude for
2.5 divisions of display on TI.
(59) Adjust ' control for best corner.
(60) Press and release upper TRIGGER COUPLING pushbutton.
39