CAL 09 Parametric Measurements

Calibration Procedure For Oscilloscope, Os-288/G (Type 2465B) And Types 2465B Opt46 And 2467B Page Navigation
  36    37    38    39    40  41
img
. . .
TB 9-6625-2295-24
Table 27. CAL 09 Period - Continued
Test instrument
Oscilloscope
CAL 09
calibrator
step numbers
output settings
6
2
Ps
7
5
Ps
8
10
Ps
9
20
Ps
10
50
Ps
11
0.1  ms
12
0.2  ms
13
0.5  ms
14
1
ms
15
2
ms
16
5
ms
17
0.2  ms
(7) TI will automatically step from step 18 through 28, and display readout will
indicate DIAGNSTIC PUSH A/B TRIG TO EXIT.
b. Adjustments. None.
24. Final Procedure
a. Press A/B TRIG pushbutton to exit, press POWER pushbutton to OFF, and move
CAL NO/CAL jumper to NO CAL position and replace protective cover.
b. Repeat paragraphs 7 through 14.