Table 26. Vertical Attenuator
CAL 06 Vertical Transient Response

Calibration Procedure For Oscilloscope, Os-288/G (Type 2465B) And Types 2465B Opt46 And 2467B Page Navigation
  31    32    33    34    35  36  37    38    39    40    41  
TB 9-6625-2295-24
b. Adjustments. None.
21. CAL 03 Triggering
a. Alignment
(1) Press upper TRIGGER MODE pushbutton to CAL 03 (bottom left of display).
(2) Press and release upper TRIGGER COUPLING pushbutton.
(3) TI will automatically step from 200 to 214 and stop at 215. Display indicates
CH 1, 500 mV, (step) 215.
(4) Connect oscilloscope calibrator output to CH 1.
(5) Set oscilloscope calibrator for a CHAN 1, VOLT mode output of .5 V at 1 kHz.
(6) Press and release upper TRIGGER COUPLING pushbutton. Display readout
will indicate CH 1, 500 mV, (step) 216.
(7) Press and release upper TRIGGER COUPLING pushbutton. Display readout
will indicate CH 2, 500 mV, (step) 217.
(8) Move connection from CH 1 to CH 2 and repeat (5) above.
(9) Press and release upper TRIGGER COUPLING pushbutton. Display readout
will indicate CH 3, 500 mV, (step) 218.
(10) Move connection from CH 2 to CH 3 and repeat (5) above.
(11) Press and release upper TRIGGER COUPLING pushbutton. Display readout
will indicate CH 3, 2V, (step) 219.
(12) Set oscilloscope calibrator for a CHAN 1, VOLT mode output of 2 V at 1 kHz.
(13) Press and release upper TRIGGER COUPLING pushbutton. Display readout
will indicate CH 4, 500 mV, (step) 220.
(14) Move connection from CH 3 to CH 4.
(15) Set oscilloscope calibrator for a CHAN 1, VOLT mode output of .5 V at 1 kHz.
(16) Press and release upper TRIGGER COUPLING pushbutton.
Display will
indicate CH 4, 2 V, (step) 221.
(17) Set oscilloscope calibrator for a CHAN 1, VOLT mode output of 2 V at 1 kHz.
(18) Press and release upper TRIGGER COUPLING pushbutton.
Display will
indicate DIAGNSTIC PUSH A/B TRIG TO EXIT.
b. Adjustments. None.
22. CAL 06 Vertical Transient Response
a. Alignment
(1) Press upper TRIGGER MODE pushbutton to CAL 06 (bottom left of display).
(2) Press and release upper TRIGGER COUPLING pushbutton. Display readout
will indicate ADJ ' (step) 1, 10 mV, 100 ns.
36