Figure 8. Example of FOCUS generated pulse equal to dead zone pulse amplitude (TD-260C).
Figure 12. Example of cursor B placement in center flat portion on FOCUS generated pulse (TD-260C).

Calibration Procedure For Optical Fiber Test Set, Ts-4320(P)/G Page Navigation
  13    14    15    16    17  18  19    20    21    22    23  
img
. . . .
TB 9-6625-2309-24
(18) Adjust CURSOR A control to place cursor A on peak (or approximate middle if
no peak is evident) of TI dead zone pulse (fig. 10) or peak of dead zone pulse if peak is evident
(TD-260C and TD-261C optical modules) (fig. 11). Record A= distance indicated on TI
display on appropriate Reference Cursor Position Data Sheet (appendix A).
Figure 10. Example of cursor A on approximate middle of dead zone pulse
with no peak evident (TD-260C shown for example).
Figure 11. Example of cursor A on peak of dead zone pulse- (TD260C).