Figure 11. Example of cursor A on peak of dead zone pulse- (TD260C).
Attenuation Scale Non-linearity - Continued

Calibration Procedure For Optical Fiber Test Set, Ts-4320(P)/G Page Navigation
  14    15    16    17    18  19  20    21    22    23    24  
img
. .
TB 9-6625-2309-24
NOTE
Take  care  not  to  move  CURSOR  A  control  during
measurements to reduce error.  If CURSOR A control is
moved, reposition cursor until TI displays A = recorded position
again. Doing this will reestablish same optical power reference
level as before and calibration may continue with no additional error.
(19) Press HOR CONT pushbutton for maximum horizontal contraction.
(20) Press DISPLAY FROM pushbutton to select B.
(21) Adjust CURSOR B control to place cursor B in approximate middle of flat
portion on top of FOCUS generated pulse (fig. 12). Record B= distance indicated on TI
display on TD-260C (850 nm) Reference Cursor Position Data Sheet (appendix A).
Figure 12. Example of cursor B placement in center flat portion on FOCUS generated pulse (TD-260C).
NOTE
Take  care  not  to  move  CURSOR  B  control  during
measurements to reduce error.  If CURSOR B control is
moved, reposition cursor B so TI displays B =  recorded
reference cursor position again. This will reestablish same
measured power level on FOCUS generated pulse as before
and calibration may continue with no additional error.
(22) Perform (a) and (b) below:
(a) Press TI LOSS MODE pushbutton to select 2 POINT.
20