Attenuation Scale Non-linearity - Continued
Figure 11. Example of cursor A on peak of dead zone pulse- (TD260C).

Calibration Procedure For Optical Fiber Test Set, Ts-4320(P)/G Page Navigation
  12    13    14    15    16  17  18    19    20    21    22  
img
. . .
TB 9-6625-2309-24
Figure 8. Example of FOCUS generated pulse equal to dead zone pulse amplitude (TD-260C).
(17) Press HOR EXP pushbutton to expand display to view dead zone pulse (fig. 9).
Figure 9. Example of expanded dead zone pulse - (TD-260C).
18