TB 9-6625-2309-24
(e) Repeat technique of steps (b) through (d) above as necessary for remaining
data on TI optical module data sheet.
(f) Verify data has been entered correctly.
(g) Press S on keyboard at prompt and enter B = reference cursor position in km.
(h) Press Q on keyboard to quit data edit page.
(42) Perform steps (a) through (c) below:
(a) Press 3 on keyboard to select Generate Report.
(b) Generate Attenuation Report menu will appear on screen.
(c) Press 2 on keyboard to select Graphic Report (Screen) then press 1 on
keyboard to select Graphic Report (Printer) to print a hard copy to be sent to customer.
Refer to note immediately following figure 14 for pass/fail condition.
NOTE
Boundaries shown on graphical report are normal nonlinearity
Distance [km]
: 0.00
Reference Attenuation [dB]
: 3.04
OFTS Distance
Scale
Pulse Width
: long
Wavelength [nm]
: 850
Model: TS-4320
Date: 8-30-94
Main Frame SN: 935-003
Optical Module SN: 112-385
23