TB 9-6625-2320-35
IDENTIFICATION AND DESCRIPTION
1. Test Instrument Identification. This bulletin
provides instructions for the
calibration of Digitizing Oscilloscope, Tektronix, Model TDS540B. The manufacturer's
manual was used as the prime data source in compiling these instructions. The
equipment being calibrated will be referred to as the TI (test instrument) throughout
this bulletin.
a. Model Variations . None.
b. Time and Technique . The time
required for this
calibration is
approximately 8
hours, using the dc and low frequency technique.
2. Forms,
Records, and
Reports
a. Forms, records, and
reports required for
calibration personnel at all
levels are
prescribed by TB 750-25.
b. Adjustments to be reported are designated (R) at the end of the sentence in which
they appear. Report only those adjustments made and designated with (R).
3. Calibration
Description. TI
parameters and
performance specifications which
pertain to this calibration are listed in table 1.
Test instrument parameters
Performance specifications
DC Gain
Range: 1 mV/div to 10 V/div
Accuracy: 1%
DC Voltage1
Accuracy: ((1.0% x | reading - Net Offset1|) + Offset Accuracy
+ (0.6 div x V/div))
Accuracy: ((1.0% x | reading|) + (0.1 div x V/div) + 0.3 mV)
Delta volts between any
two averages of ≥16
the same setup and
ambient conditions
DC to 500 MHz
Range: 10 mV/div to 1V/div
5 mv/div to 9.95 mV/div
2 mv/div to 4.98 mV/div
DC to 500 MHz
DC to 450 MHz
Range: 1 mv/div to 1.99 mV/div
See footnotes at end of table.