TB 9-6625-2320-35
(6) Press side-menu OK Confirm Run Test key.
NOTE
This verification will take up to three and a half minutes. A
"clock" icon will appear on-screen during the verification and
disappear when verification is complete.
(7) When the verification is complete, verify that no failures are found and
reported on-screen.
(8) Press front-panel SHIFT key.
(9) Press front-panel UTILITY key.
(10) Repeatedly press main-menu System <config> key until <Cal> is
highlighted in the pop-up menu.
(11) Verify that the word Pass appears in the main-menu under the following
menu labels: Voltage Reference , Frequency Response, and Pulse Trigger. See
7