(12) Press main-menu Signal Path key.
(13) Press side-menu OK Compensate Signal Paths key.
NOTE
This verification will take up to five minutes. A "clock" icon
will appear on-screen during the verification and disappear
when verification is complete.
(14) When the verification is complete, verify that the word Pass appears under
Signal Path in the main menu.
(15) Press front panel CLEAR MENU key.
NOTE
When indications specified in paragraphs b through d are
Process.
b. Input Channel Functional Test
(1) Install probe on CH 1 .
(2) Connect probe tip to front-panel PROBE COMPENSATION SIGNAL;
connect probe ground to front-panel PROBE COMPENSATION GND.
(3) Press front-panel SETUP key.
(4) Press main-menu Recall Factory Setup key.
(5) Press side-menu OK Confirm Factory Init key.
NOTE
This verification will take up to 15 seconds. A "clock" icon
will appear on-screen during the verification and disappear
when verification is complete.
(6) Press front-panel WAVEFORM OFF key.
(7) Press front-panel CH 1 key.
(8) Press front-panel TRIGGER MENU key.
(9) Press main-menu Source key.
(10) Press side-menu CH1 key.
8