TB 9-6625-2320-35
(9) Press main-menu Mode & Holdoff key.
(10) Press side-menu Normal key.
(11) Press front-panel CLEAR MENU key.
(12) Verify front-panel TRIGGER MAIN LEVEL knob can trigger and untrigger
the squarewave signal as it is rotated. Leave the signal untiggered, which is indicated
by the display not updating.
(13) Press front-panel SET LEVEL TO 50% key and verify the squarewave signal
triggers. Leave the signal triggered.
(14) Press front-panel HORIZONTAL MENU key.
(15) Press main-menu Time Base key.
(16) Press side-menu Delayed Triggerable key.
(17) Press side-menu Delayed Only key.
(19) Adjust front-panel HORIZONTAL SCALE knob to display 200 s.
(20) Press front-panel SHIFT key.
(21) Press front-panel DELAYED TRIG key.
(22) Press main-menu Level key.
(23) Press side-menu Level key.
(24) Verify delayed trigger level readout changes as front-panel general purpose
knob is rotated.
(25) Verify squarewave signal triggers and untriggers as front-panel general
purpose knob is rotated. Leave signal untriggered, which is indicated by the front-panel
"READY" display not updating.
(26) Press side-menu Set to 50% key and verify probe-compensation signal
triggers. Leave signal triggered.
(27) Press main-menu Delay by Time key.
(28) Press front-panel 1 key.
(29) Press front-panel ENTER key.