TB 9-6625-2320-35
(22) Press front-panel ENTER key.
readout does not indicate between 1.0355 and 1.0445 V, perform b below.
(24) Repeat technique of (16) through (23) above, reversing the polarity of the
position, offset, and generator settings. If TI C1 Mean readout does not indicate
between -1.0445 and -1.0355 V, perform b below.
(25) Repeat technique of (15) through (24) above for settings listed in table 4. If TI
does not indicate within limits listed in table 4, perform b below.
Test instrument
Test instrument
Test
Test instrument
VERTICAL
Position
instrument
generator
C1 Mean
SCALE
settings
Offset
output
indications
settings
(div)
settings (V)
settings
Min
Max
200 mV
-5
+10
+11.6 V
+11.5195 V
+11.6805
V
+5
-10
-11.6 V
-11.6805 V
-11.5195
V
1V
-5
+10
+18 V
+17.7575 V
+18.2425
V
+5
-10
-18 V
-18.2425 V
-17.7575
V
(26) Press WAVEFORM OFF key to remove current channel displayed.
(27) Repeat technique of (9) through (25) above for channels 2, 3, and 4. If TI does
not indicate within limits listed in table 4, perform b below.
a. Performance Check
(1) Connect leveled sine wave generator leveling head to CH 1 .
(2) Press front-panel SETUP key.
(3) Press main-menu Recall Factory Setup key.
(4) Press side-menu OK Confirm Factory Init key.
(5) Press front-panel TRIGGER MENU key.
(6) Press main-menu Coupling key.
(7) Press side-menu Noise Rej key.