TB 9-6625-2320-35
NOTE
As you are positioning the leveled sine wave generator
amplitude, press SET LEVEL TO 50% frequently to speed
up the updating of the waveform amplitude.
(10) Position leveled sine wave generator controls for a 250 MHz output and
amplitude for about six divisions on CH 1 .
(11) Press front-panel ZOOM key.
(12) Press side-menu On key.
(13) Adjust front-panel HORIZONTAL SCALE knob to display 250 ps .
(14) Verify vertical scale factor is 1.0X and horizontal scale factor is 2.0X.
NOTE
If TI Vert and Horz readouts do not indicate as specified,
adjust front-panel HORIZONTAL SCALE knob and
VERTICAL SCALE knob.
(15) Press CH 2 key.
(16) Verify vertical scale factor is 1.0X.
17