TB 9-6625-2320-35
(8) Adjust front-panel HORIZONTAL SCALE knob to display 50 ns .
(9) Press front-panel SHIFT key.
(10) Press front-panel ACQUIRE MENU key.
(11) Press main-menu Mode key.
(12) Press side-menu Average 16 key.
(13) Press front-panel MEASURE key.
(14) Press main-menu High-Low Setup key.
(15) Press side-menu Min-Max key.
(16) Position leveled sine wave generator controls for a 6 MHz output.
(17) Press front-panel TRIGGER MENU key.
(18) Press main-menu Source key.
(19) Press side-menu CH 1 key.
(20) Press front-panel VERTICAL MENU key.
(21) Press main-menu Coupling key.
(22) Press side-menu Ω key to select 50 Ω setting.
(23) Adjust front-panel VERTICAL SCALE knob to display 100 mV.
(24) Press front-panel MEASURE key.
(25) Press main-menu Select Measrmnt for CH1 key.
(26) Press side-menu more key until the menu label Frequency appears in the
side-menu.
(28) Press side-menu more key until the menu label Pk-Pk appears in the side-
menu.
(29) Press side-menu Pk-Pk key.
(30) press front-panel CLEAR MENU key.