TB 9-6625-2320-35
a. Performance Check
(1) Connect calibrator OUTPUT to CH1.
(2) Press front-panel SETUP key.
(3) Press main-menu Recall Factory Setup key.
(4) Press side-menu OK Confirm Factory Init key.
(5) Press front-panel SHIFT key.
(6) Press front-panel ACQUIRE MENU key.
(7) Press main-menu Mode key.
(8) Press side-menu Average 16 key.
(9) Press front-panel CH 1 key.
(10) Press front-panel MEASURE key.
(11) Press main-menu Select Measrmnt for Ch 1 key.
(12) Press side-menu more key until the menu label Mean appears in the side-
menu.
(13) Press side-menu Mean key.
(14) Press front-panel CLEAR MENU key.
(15) Adjust front-panel VERTICAL SCALE knob to display 5 mV.
(16) Press front-panel VERTICAL MENU key.
(17) Press main-menu Position key.
(18) Use front-panel keypad to set vertical position to -5 divisions. Press front-
panel - and 5 keys.
(19) Press front-panel ENTER key.
(20) Press main-menu Offset key.
(21) Use front-panel keypad to set vertical offset to 1 volt. Press front-panel 1
key.