TB 9-6625-2320-35
(17) Press front-panel HORIZONTAL MENU key.
(18) Press main-menu Time Base key.
(19) Press side-menu Delayed Only key.
(20) Press side-menu Delayed Runs After Menu key.
(21) Adjust front-panel HORIZONTAL SCALE delayed time base (D) to 100 ns.
(22) Set delayed time to 10 ms. Press front-panel 1 and 0 keys.
(23) Press front-panel SHIFT key.
(24) Press front-panel m key.
(25) Press front-panel ENTER key.
(26) If TI rising edge of the marker does not cross the center horizontal graticule
line at a point within 2.5 divisions of center graticule, perform b below.
14. Pulse-Width Triggering
a. Performance Check
(1) Press front-panel SETUP key.
(2) Press main-menu Recall Factory Setup key.
(3) Press side-menu OK Confirm Factory Init key.
(4) Press front-panel VERTICAL MENU key.
(5) Press main-menu Coupling key.
(6) Press side-menu Ω key to select 50 Ω setting.
(7) Adjust front-panel HORIZONTAL SCALE knob to display 12.5 ns .
(8) Connect leveled sine wave generator leveling head to CH 1 using X10
attenuator.
(9) Position leveled sine wave generator controls for a 100 MHz output and five
division sine wave.