TB 9-6625-2320-35
(29) Press side-menu Lower Limit key.
(30) Rotate general purpose knob to increase the Lower Limit readout until
triggering is lost.
(31) If TI Lower Limit readout, after TI loses triggering, does not indicate within
3.5 ns to 6.5 ns, perform b below.
(32) Repeat (23) through (27) above to return lower limits to 2 ns.
(33) Press side-menu Upper Limit key.
(34) Rotate general purpose knob to decrease the Upper Limit readout until
triggering is lost.
(35) If TI Upper Limit readout, after TI loses triggering, does not indicate within
3.5 ns to 6.5 ns, perform b below.
(36) Press side-menu Upper Limit key.
(37) Set upper limit to 4 s. Press front-panel 4 key.
(38) Press front-panel SHIFT key.
(39) Press front-panel key.
(40) Press front-panel ENTER key.
(41) Press side-menu Lower Limit key.
(42) Set lower limit to 500 ns. Press front-panel 5, and 0 (twice) keys.
(43) Press front-panel SHIFT key.
(44) Press front-panel n key.
(45) Press front-panel ENTER key.
(46) Adjust front-panel HORIZONTAL SCALE knob to display 5 s.
(47) Position leveled sine wave generator controls for a 250 kHz output, five
division sine wave.
(48) Adjust front-panel VERTICAL SCALE knob to display 20 mV.