TB 9-6625-2320-35
(4) Press main-menu Recall Factory Setup key.
(5) Press side-menu OK Confirm Factory Init key.
(6) Adjust front-panel VERTICAL SCALE knob to display 200 mV.
(7) Press front-panel VERTICAL MENU key.
(8) Press main-menu Position key.
(9) Set vertical position to -3 divisions. Press front-panel - and 3 keys.
(10) Press front-panel ENTER key.
(11) Press main-menu Offset key.
(12) Set vertical offset to +10 volts. Press front-panel 1 and 0 keys.
(13) Press front-panel ENTER key.
(14) Position calibrator for a +10 V output.
(15) Press front-panel SET LEVEL TO 50% key.
(16) Press front-panel TRIGGER MENU key.
(17) If TI main-menu Level readout does not indicate between 9.9393 and 10.1147
V, perform b below.
(18) Press main-menu Slope key.
(19) Press side-menu negative slope
key.
(20) Repeat (14) and (15) above.
(21) If TI Level readout does not indicate between 9.9393 and 10.1147 V, perform
b below.
(22) Press main-menu HORIZONTAL MENU key.
(23) Press main-menu Time Base key.
(24) Press side-menu Delayed Only key.
(25) Press side-menu Delayed Triggerable key.
25