(52) Press front-panel HORIZONTAL MENU key.
(53) Press main-menu Time Base key.
(54) Press side-menu Main Only key.
(55) Press front-panel CLEAR MENU key.
(56) Remove X10 attenuator and reconnect to CH 1 .
(57) Position leveled sine wave generator controls for an amplitude of 2.5
divisions.
(58) Position leveled sine wave generator controls for a 250 mV C1 Ampl readout.
Readout may fluctuate.
(59) Press front-panel TRIGGER MENU key.
(60) Press main-menu Source key.
(61) Press side-menu -more- key until DC Aux appears in the side-menu.
(62) Press side-menu DC Aux key.
(63) Press front-panel SET LEVEL TO 50% key.
(64) Press main-menu Slope key.
(65) If TI waveform trigger is not stable for both the positive and negative slopes,
perform b below. Press side-menu
and
keys to switch between trigger slopes.
Adjust front-panel TRIGGER LEVEL knob to stabilize trigger if required.
(66) Press side-menu
key.
(67) Press main-menu Source key.
(68) Press side-menu -more- key until CH 1 appears in the side-menu.
(69) Press side-menu Ch1 key.
head to CH 1 .
(71) Adjust front-panel HORIZONTAL SCALE knob to display a main (M) time
base of 500 ps.