TB 9-6625-2320-35
(8) Press front-panel SHIFT key.
(9) Adjust front-panel ACQUIRE MENU key.
(10) Press main-menu Mode key.
(11) Press side-menu Average key.
(12) Adjust front-panel general purpose knob to set averages to 64.
(13) Press front-panel WAVEFORM OFF key.
(14) Press front-panel CH 2 key.
(15) Adjust front-panel VERTICAL SCALE knob to display 1 V.
(16) Adjust front-panel VERTICAL POSITION knob to center test signal.
(17) Press front-panel MEASURE key.
(18) Press main-menu Select Measrmnt for Ch2 key.
(19) Repeatedly press side-menu -more- key until High and Low appear in side-
menu.
(20) Press side-menu High and Low keys.
(21) If TI C2 High readout does not indicate ≥2.5 V and C2 Low does not indicate
≤0.7 V, perform b below.
(22) Press front-panel VERTICAL MENU key.
(23) Press main-menu Coupling key.
(24) Press side-menu Ω key to select 50 Ω setting.
(25) If TI C2 High readout does not indicate ≥1.0 V and C2 Low does not indicate
≤0.25 V, perform b below.
(26) Remove connection from rear-panel Main Trigger Output and connect rear-
panel Delayed Trigger Output to CH 2 .
(27) If TI C2 High readout does not indicate ≥1.0 V and C2 Low does not indicate
≤0.25 V, perform b below.
(28) Press side-menu Ω key to select 1 MΩ setting.
32