TB 9-6625-2320-35
NOTE
When ON/STBY button is pressed, the TI automatically
performs a self-test sequence. Upon successful completion of
self-test, TI will be in normal operating mode.
b. Press ON/STBY button and allow at least 20 minutes for equipment warm-up.
NOTE
Steps a through g are instructions required to initially
install the TDS 500B and 700A Field Adjustment Software
and will not need to be performed again unless the software
is corrupted. If software is already installed, proceed to note
prior to step g and continue adjustment process.
a. Install a working copy disk of the TDS 500B and 700A Field Adjustment Software
in the electronic technical bulletin (ETB) process controller (PC) floppy disk drive and
type install at the floppy disk drive prompt.
b. The installation program will prompt you to specify information. Respond as
follows:
(1) Hard Drive: Enter C or drive of your choice.
(2) Computer Type: Enter 2.
(3) GBIP Board Type: Enter 1.
(4) DMA Channel: Enter 0.
(5) Interrupt: Enter 0.
(6) Switch Setting: Enter 68.
c. The installation program will create a directory on the hard drive called
TDS700.ADJ.
37