TB 9-6625-2320-35
(29) Press front-panel CLEAR MENU key.
(30) If TI C2 High readout does not indicate ≥2.5 V and C2 Low does not indicate
≤0.7 V, perform b below.
(31) Remove connection from rear-panel Delayed Trigger Output and connect
rear-panel SIGNAL OUT to CH 2 .
(32) Press front-panel TRIGGER MENU key.
(33) Press main-menu Source key.
(34) Press side-menu Ch3 key.
(35) Adjust front-panel VERTICAL SCALE knob to display 100 mV.
(36) Press front-panel SET LEVEL TO 50% key.
(37) Press front-panel MEASURE key.
(38) Press main-menu Select Measrmnt for Ch2 key.
(39) Repeatedly press side-menu -more- key until Pk-Pk appears in side-menu.
(40) Press side-menu Pk-Pk key.
(41) Press front-panel CLEAR MENU key.
(42) If TI Ch2 Pk-Pk readout does not indicate between 88 mV and 132 mV,
perform b below.
(43) Press front-panel VERTICAL MENU key.
(44) Press main-menu Coupling key.
(45) Press side-menu Ω key to select 50 Ω setting.
(46) Press front-panel CLEAR MENU key.
(47) If TI Ch2 Pk-Pk readout does not indicate between 44 mV and 66 mV,
perform b below.
33