(72) Press front-panel HORIZONTAL MENU key.
(73) Press main-menu Time Base key.
(74) Press side-menu Delayed Only key.
(75) Press side-menu Delayed Triggerable key.
(76) Adjust front-panel HORIZONTAL SCALE knob to display a delayed (D) time
base of 500 ps.
(77) Press side-menu Main Only key.
(78) Position leveled sine wave generator controls for a 500 MHz output, five
division sine wave.
(79) Position leveled sine wave generator controls for a 500 mV C1 Ampl readout.
Readout may fluctuate. Press front-panel SET LEVEL TO 50% key to stabilize readout
if required.
(80) Disconnect leveling head from CH 1 and reconnect it to CH 1 through an X5
attenuator.
(81) Repeat technique of (34) through (50) above.
17. Trigger Output Signal
a. Performance Check
to CH 3 .
(2) Position calibration generator controls for 0.500 V, 100Hz output.
(3) Connect rear-panel Main Trigger Out to CH 2 .
(4) Press front-panel SETUP key.
(5) Press main-menu Recall Factory Setup key.
(6) Press side-menu OK Confirm Factory Init key.
(7) Adjust front-panel HORIZONTAL SCALE knob to display 200 s.