TB 9-6625-2320-35
Test instrument parameters
Performance specifications
Offset
Range: 1 mV/div to 100mV/div
Accuracy: ((0.2% x | Net Offset1 |) + 1.5 mV + (0.1 div x V/div))
Range: 101 mV/div to 1 V/div
Accuracy: ((0.25% x | Net Offset1 |) + 15 mV + (0.1 div x V/div))
Range: 1.01 V/div to 10 V/div
Accuracy: ((0.25% x | Net Offset1 |) + 150 mV + (0.1 div x
V/div))
≤50 ps for any two channels with equal volts/div and coupling
Delay
settings
100 ppm over any ≥1 ms interval3
Time Base
Range: Dc to 1 MΩ Coupled
Accuracy: 1 MΩ 0.5% in parallel with 10 pF 3 pF
Range: DC to 50 MΩ Coupled
Accuracy: 50 Ω 1% with VSWR ≤1.3:1 from Dc to 500 MHz,
≤1.5:1 from 500 MHz to 1 GHz
Range: DC - 1 MΩ, AC - 1 MΩ, or GND Coupled
Input Voltage
Accuracy: 300 V (DC + peak AC), 400 V peak; derate at
20 dB/decade above 1 MHz, CAT II
Range: Dc - 50Ω or ac - 50Ω coupled
Accuracy: 5 VRMS, with peaks ≤+30 V
Accuracy: 25 ppm over any ≥1 ms interval
Long Term Sample Rate and
Delay Time
Sensitivity and Edge Trigger
Any Channel:
0.35 div from dc to 50 MHz, increasing to 1 div at
500 MHz
Auxiliary : 400 mV from dc to 50 MHz, increasing to 750 mV at
100 MHz
Range: 1 ns to 1 s
Pulse-Width Triggering
Accuracy: (20% of setting +0.5 ns)
Range: 1.02 s to 1 s
Accuracy: (100 ns + 0.01% of setting)
Trigger Output
VOUT(HI)
Accuracy: ≥2.5 open circuit; ≥1.0 V into a 50 Ω load to ground
Vout (LO)
Accuracy: ≤0.7 V into a load of ≤4 mA; ≤0.25 V into a 50 Ω load to
ground
1Net
Offset = Offset - (Position x Volts/Div). Net Offset is the nominal voltage level at the oscilloscope input
that corresponds to the center of the A-D converter dynamic range. Offset Accuracy is the accuracy of this
voltage level
2The limits given are for the ambient temperature range of 0C to +30C. Reduce the upper bandwidth
frequencies by 2.5 MHz for each C above +30C
3Due to STANDARDS limitations, this specification does not agree with manufacturer's stated accuracies.
EQUIPMENT REQUIREMENTS
4. Equipment
Required. Table 2 identifies the
specific equipment to be
used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-286. Alternate items may be used by the calibrating activity.
The items selected must be verified to perform satisfactorily prior to use and must bear