Figure 1. Time base verification hook-up.
Table 3. Frequency Resolution Accuracy

Calibration Procedure For Signal Generator Model 68347M Page Navigation
  1    2    3    4    5  6  7    8    9    10    11  
img
. . . .
TB 9-6625-2322-24
Figure 3. Frequency resolution accuracy hook-up.
(2) Press TI keys as listed in (a) through (e) below.
SYSTEM, Reset.
(a)
OUTPUT off.
(b)
Level Control, L0, Edit L0, 0, and dB.
(c)
Frequency Control, F0, Edit F0, 1, and GHz.
(d)
OUTPUT on.
(e)
(3) Set microwave frequency counter frequency on INPUT 1.
(4) Verify that the microwave frequency counter indicates within limits listed
in table 3.
(5) Set TI frequency to the next frequency listed in table 3 using the TI arrow pad
and repeat (4) above.
(6) Repeat (4) and (5) above for the remaining frequencies listed in table 3.
7