Table 5. 2 GHZ Output Level Test 1 dB Steps
Figure 4. Spectral purity hook-up.

Calibration Procedure For Signal Generator Model 68347M Page Navigation
  5    6    7    8    9  10  11    12    13    14    15  
img
. . . . .
TB 9-6625-2322-24
Table 6. 2 GHz Output Level Test 10 dB Steps
Test instrument
Measuring
Measuring
output level
receiver
receiver
(dB)
indication
indication
Min (dB)
Max (dB)
0
-1
1
-10
-11
-9
-20
-21
-19
-30
-31
-29
-40
-41
-39
-50
-51
-49
-60
-61
-59
-70
-71
-69
-80
-81
-79
-90
-91
-89
(11) Press TI keys as listed in (a) and (b) below:
(a) Frequency Control, F0, Edit F0, 5, and GHz.
(b) Level Control, L0, Edit L0, 0, and dB.
(12) Using measuring receiver and RF power techniques in Log Mode verify that the
measuring receiver indicates within tolerances listed in table 7.
(13) Using TI arrow key set the TI to the remaining levels listed in table 7 and repeat
(12) above.
Table 7. 5 GHz Output Level Test 1 dB Steps
Test instrument
Measuring
Measuring
output level
receiver indication  receiver indication
(dB)
Min (dB)
Max (dB)
0
-1
1
-1
-2
0
1
0
2
2
1
3
3
2
4
4
3
5
5
4
6
6
5
7
7
6
8
8
7
9
9
8
10
10
9
11
11
10
12
(14) Press Level Control, L0, Edit L0, 0, and dB.
(15) Using standard tuned level measurement techniques verify the measuring
receiver indicates within minimum and maximum limits for TI output level as listed in
table 8 below.
11