Table 9. Spectral Purity
Amplitude Modulation

Calibration Procedure For Signal Generator Model 68347M Page Navigation
  8    9    10    11    12  13  14    15    16    17    18  
img
. . . .
TB 9-6625-2322-24
Table 9. Spectral Purity - Continued
Test instrument
Spectrum analyzer
Harmonic
Maximum
frequency
reference frequency
frequency
indication
2.1  GHz
2.1  GHz
14.7  GHz
-60
2.1  GHz
2.1  GHz
16.8  GHz
-60
3.6  GHz
3.6  GHz
----------
13
3.6  GHz
3.6  GHz
7.2  GHz
-60
3.6  GHz
3.6  GHz
10.8  GHz
-60
3.6  GHz
3.6  GHz
14.4  GHz
-60
3.6  GHz
3.6  GHz
18.0  GHz
-60
7.0  GHz
7.0  GHz
----------
13
7.0  GHz
7.0  GHz
14.0  GHz
-60
(5) Reduce all outputs to minimum.
(6) Disconnect TI from spectrum analyzer.
b. Adjustments. No further adjustments can be made.
12. Amplitude Modulation
a. Performance Check
(1) Connect equipment as shown in figure 5.
Figure 5. Amplitude modulation hook-up.
14