Amplitude Modulation - Continued
Table 12. FM Deviation Locked and Response

Calibration Procedure For Signal Generator Model 68347M Page Navigation
  11    12    13    14    15  16  17    18    19    20    21  
img
. . . .
TB 9-6625-2322-24
(15) Reduce all outputs to minimum.
(16) Disconnect equipment setup.
b. Adjustments. No further adjustments can be made.
13. Frequency Modulation
a. Performance Check
(1) Connect equipment as shown in figure 6 below.
Figure 6. Frequency modulation hook-up.
(2) Set measuring receiver to measure frequency modulation, with +PEAK detector,
<20 Hz high pass and >200 kHz lo pass filters.
(3) Press TI keys as listed in (a) through (j) below.
(a) SYSTEM, Rest.
(b) OUTPUT off.
(c) Level Control, L0, Edit L0, 0, and dB.