Table 6. 2 GHz Output Level Test 10 dB Steps
Table 9. Spectral Purity

Calibration Procedure For Signal Generator Model 68347M Page Navigation
  6    7    8    9    10  11  12    13    14    15    16  
img
. . . . . . .
TB 9-6625-2322-24
(16) Use the TI arrow key to decrement the output level 10 dB as indicated in table 8
and repeat (15) above.
(17) Repeat (16) above for remaining TI output level settings listed in table 8.
Table 8. 5 GHz Output Level Test 10 dB Steps
Test instrument
Measuring
Measuring
output level
receiver
receiver
(dB)
indication
indication
Min (dB)
Max (dB)
0
-1
1
-10
-11
-9
-20
-21
-19
-30
-31
-29
-40
-41
-39
-50
-51
-49
-60
-61
-59
-70
-71
-69
-80
-81
-79
-90
-91
-89
(18) Reduce TI output to minimum.
(19) Disconnect TI from measuring receiver.
b. Adjustments. No adjustments can be made.
11. Spectral Purity
a. Performance Check
(1) Connect equipment as shown in figure 4.
Figure 4. Spectral purity hook-up.