TB 9-6625-2327-24
Test instrument settings
Multimeter indications (V)
Amplitude (Vrms)
Min
Max
(Hz)
10
m
1
k
9.546
10.454
100
m
1
k
98.646
101.354
200
m
1
k
0.197646
0.202354
300
m
1
k
0.296646
0.303354
400
m
1
k
0.395646
0.404354
500
m
1
k
0.494646
0.505354
600
m
1
k
0.593646
0.606354
1
670.0
m
1
k
0.662946
0.677054
2
670.0
m
100
k
0.663300
0.676700
700
m
1
k
0.692646
0.707354
800
m
1
k
0.791646
0.808354
900
m
1
k
0.890646
0.909354
1.000
1
k
0.989646
1.010354
3
2.000
1
k
1.979646
2.020354
4
2.000
100
k
1.980000
2.020000
5
7.000
1
k
6.929646
7.070354
6
7.000
100
k
6.930000
7.070000
Record multimeter indication as 1kHz_0dB_reference in table 5.
1
Record multimeter indication as 100kHz_0dB_reference in table 5.
2
3 Record multimeter indication as 1kHz_10dB_reference in table 5.
4 Record multimeter indication as 100kHz_10dB_reference in table 5.
5 Record multimeter indication as 1kHz_20dB_reference in table 5.
6 Record multimeter indication as 100kHz_20dB_reference in table 5.
(6) Set Output to off and disconnect equipment.
b. Adjustments. Perform entire alignment procedure listed in paragraph 13.
a. Performance Check
Multimeter data
Formula to calculate
Calculated dB
Reference
dB reference
reference
1kHz_0dB_reference
100kHz_0dB_reference
10 * Log(5.0 * multimeter data2)
1kHz_10dB_reference
10 * Log(5.0 * multimeter data2)
100kHz_10dB_reference
10 * Log(5.0 * multimeter data2)
1kHz_20dB_reference
10 * Log(5.0 * multimeter data2)
100kHz_20dB_reference
(2) Connect power sensor module to power meter, zero and calibrate.
(3) Connect power sensor to TI Output connector.