TB 9-6625-2329-24
Table 1. Calibration Description - Continued
Test instrument parameters
Performance specifications
Frequency range: 10 Hz to 100 kHz
Accuracy: +40 PPM
Level range: 2 mV to 2 V
Accuracy: +1% + 1 mV
Frequency response range:
Accuracy:
+2.5%
10 Hz to 20 kHz
20 Hz to 100 kHz
+3.5%
Distortion: >0.5 V level
Accuracy: <0.1% (10 Hz to 100 kHz)
1Verified
to -110 dB.
Verified from 30 Hz to 50 kHz.
2
2Verified to 400 kHz dev.
4Verified from 30 Hz to 50 kHz.
5Verified from 300 Hz to 5 kHz.
EQUIPMENT REQUIREMENTS
4. Equipment Required. Table 2 identifies the specific equipment to be used in this
calibration procedure. This equipment is issued with Secondary Transfer Calibration
Standards Set AN/GSM-286, AN/GSM-287 and AN/GSM-705. Alternate items may be used
by the calibrating activity. The items selected must be verified to perform satisfactorily
prior to use and must bear evidence of current calibration. The equipment must meet or
exceed the minimum use specifications listed in table 2. The accuracies listed in table 2
provided a four-to-one ratio between the standard and TI. Where the four-to-one ratio
cannot be met, the actual accuracy of the equipment is shown in parenthesis.
5. Accessories Required. The accessories required for the calibration are common usage
accessories, issued as indicated in paragraph 4 above, and are not listed in this calibration
procedure.
Table 2. Minimum Specifications of Equipment Required
Manufacturer and model
Common name
Minimum use specifications
(part number)
AUDIO ANALYZER
Frequency: 10 Hz to 100 kHz
Boonton, Model 1121 (1121)
Distortion capability:
<0.1%
Fluke, Model PM6681/656
FREQUENCY
Frequency: 10 Hz to1000 MHz
10
(PM 6681/656)
COUNTER
Accuracy:
+1.25 part in 10
9
FREQUENCY
Range capability: 1 part in 10
Tracor, Model 527E
DIFFERENCE METER
MEASURING
Power measurement:
Measuring receiver system
RECEIVER
N5531S consisting of:
Spectrum Analyzer, Agilent
Level: +13 dB to -110 dB
Model E4440A (E4440A),
Accuracy: +0.375 dB
Power meter, Agilent Model
Flatness measurement:
E4419B (E4419B), and Sensor
module, Agilent Model 504
Level: 0 dBm
Accuracy: +0.25 dB