SECTION II. EQUIPMENT REQUIREMENTS
Figure 1. Internal oscillator stability.

Calibration Procedure For Signal Generator Model 801.0001.43 Page Navigation
  1    2    3  4  5    6    7    8    9  
img
. . . . . . . . .
TB 6-6625-2329-24
Table 2. Minimum Specifications of Equipment Required - Continued
Manufacturer and model
Common name
Minimum use specifications
(part number)
MULTIMETER
Voltage measurement:
Hewlett-Packard, Model 3458A
Frequency: 10 Hz to 100 kHz
(3458A)
Level: 2 V to 2 mV ac
Accuracy: 0.25%
PULSE GENERATOR
Amplitude:  5 V
LeCroy, Model 9210 (9210)
Period: 10 ms
w/plug-ins, LeCroy Models
Width: 5 ms
9211 (9211) and 9215 (9215)
(MIS 45839)
SPECTRUM ANALYZER
(AN/USM-677)
Harmonics measurement:
Frequency: 100 kHz to 1000 MHz
Level: 13 to -80 dB
Accuracy: <-80 dB
TIME/FREQUENCY
Frequency: 1 MHz
Datum, Model ET6000-75
Accuracy:  2.5 part in 10
11
WORKSTATION
(13589305)
SECTION III
CALIBRATION PROCESS
6. Preliminary Instructions
a. The instructions outlined in paragraphs 6 and 7 are preparatory to the calibration
process. Personnel should become familiar with the entire bulletin before beginning the
calibration.
b. Items of equipment used in this procedure are referenced within the text by common
name as listed in table 2.
c. Unless otherwise specified, verify the result of each test and, whenever the test
requirement is not met, take corrective action before continuing with the calibration.
Adjustments required to calibrate the TI are included in this procedure.  Additional
maintenance information is contained in manufacturer's manual for this TI.
d. When indications specified in paragraphs 7 through 16 are not within tolerance,
perform the power supply check prior to making adjustments. After adjustments are made,
repeat paragraphs 7 through 16.  Do not perform power supply check if all other
parameters are within tolerance.
e. Unless otherwise specified, all controls and control settings refer to the TI.
7. Equipment Setup
WARNING
HIGH VOLTAGE is used or exposed during the performance of
this calibration.
DEATH ON CONTACT may result if
personnel fail to observe safety precautions.
REDUCE
OUTPUT(S)  to  minimum  after  each  step  within  the
performance check.
5