TB 9-6625-2329-24
Test instrument
Frequency counter indications
DATA and
(MHz)
ENTER/UNITS
(MHz)
Min
Max
10
9.999999
10.000001
60
59.999999
60.000001
90
89.999999
90.000001
150
149.999999
150.000001
(5) Press PARAMETER LEVEL pushbutton and press ON/OFF OFF key.
(6) Disconnect TI RF 50
output to frequency counter C input.
(7) Connect TI RF 50
(9) Press PARAMETER LEVEL pushbutton and press ON/OFF INT ON key.
(10) Press TI PARAMETER RF, enter DATA numerical value, and ENTER/UNITS
MHz key for each row in table 4. Frequency will indicate within limits specified.
Test instrument
Frequency counter indications
(MHz)
DATA and
ENTER/UNITS
Min
Max
(MHz)
400
399.999999
400.000001
700
699.999999
700.000001
1000
999.999999
1000.000001
(11) Press PARAMETER LEVEL pushbutton and press ON/OFF OFF key.
(12) Disconnect TI RF 50
from frequency counter.
b. Adjustments. No adjustments can be made.
9. RF Output and Flatness
a. Performance Check
NOTE
Zero, calibrate, and save sensor values as necessary
(1) Set measuring receiver to measure RF power:
(2) Connect sensor module input to TI RF 50 .
(3) Press pushbuttons as listed in (a) through (d) below:
(a) SHIFT INSTR PRESET.
(b) PARAMETER RF.
(c) DATA-100.
(d) ENTER/UNITS MHz.