TB 6-6625-2329-24
(5) Press TI PARAMETERRF pushbutton, enter DATA 8 numerical value, and
press ENTER/UNITS MHz key.
(6) Using measuring receiver RF power measurement techniques, measure RF
power and set RATIO on.
(7) Press TI PARAMETERRF pushbutton, enter DATA 7.999 numerical value
and press ENTER/UNITS MHz key.
(8) Set measuring receiver to measure RF power at 7.999 Mhz.
(9) Adjust R641 (fig. 4) for a 0.00 0.05 dBm power indication on measuring receiver. (R)
(10) Press SHIFT-INSTR PRESET
PARAMETERLEVEL
pushbuttons,
pushbutton, and press ON/OFFOFF key.
(11) Replace top cover.
a. Performance Check
(1) Ensure measuring receiver, sensor module input is connected to TI RF 50 .
NOTE
On measuring receiver zero, calibrate, and save sensor values
as necessary.
(2) Using measuring receiver with sensor module, press PRESET pushbutton.
(3) Press pushbuttons as listed in (a) through (d) below:
SHIFT-INSTR PRESET.
(a)
PARAMETER-RF.
(b)
DATA-30.
(c)
ENTER/UNTS-MHz.
(d)
(4) Setup measuring receiver to measure RF power.
(5) Press TI PARAMETERLEVEL pushbutton, enter DATA numerical value
and press ENTER/UNITS dBm key for each row in table 7. Using measuring receiver a n d
RF power measurement techniques, measured power will indicate within limits
specified.