Table 6. Output Level Flatness
Figure 4. Adjustment locations.

Calibration Procedure For Signal Generator Model 801.0001.43 Page Navigation
  4    5    6    7    8  9  10    11    12    13    14  
img
.
TB 9-6625-2329-24
Table 6. Output Level Flatness - Continued
Measuring receiver
Test instrument
indications
DATA and
ENTER/UNITS
MHz
Reading
Maximum limit
(dB)
<(dB)
7
1
8
1
9
1
10
1
20
1
30
1
40
1
50
1
60
1
70
1
80
1
90
1
100
1
200
1
300
1
400
1
500
1
600
1
700
1
800
1
900
1
1000
1
b. Adjustments
(1) Press  SHIFT INSTR  PRESET
PARAMETERLEVEL
pushbuttons,
pushbutton, then press ON/OFF-OFF key.
(2) Remove top cover.
(3) Press pushbuttons as listed in (a) through (f) below:
PARAMETER RF.
(a)
DATA 100.
(b)
ENTER/UNITSMHz.
(c)
PARAMETERLEVEL.
(d)
DATA0.
(e)
ENTER/UNITS-dBm.
(f)
(4) Adjust R514 (fig 4) for a 0 0.1 dBm power indication (R).