Table 18. Frequency Response
Residual Responses

Calibration Procedure For Spectrum Analyzer Models 8562Ec And 8562Ec-104 Page Navigation
  26    27    28    29    30  31  32    33  
img
. . . . .
TB 9-6625-2333-24
(22) Adjust synthesized signal generator level output for a TI MKR amplitude
indication between -9.95 and -10.05 dBm. Measuring receiver indication will be within
r1.8 dB of indication recorded in (7) above.
(23) Record measuring receiver indication.
(24) Connect equipment as shown in figure 6.
Figure 6. Frequency response - equipment setup.
(25) Press synthesizer/level generator keys to values as listed in (a) through (c) below:
(a) FREQUENCY to 10 MHz.
(b) AMPLITUDE to -4 dBm.
(c) AMPTD INCR to 0.1 dB.
(26) Press TI MARKER PEAK SEARCH key.
(27) Slowly adjust synthesizer/level generator amplitude until TI MKR indication is
as close as possible to measuring receiver indication recorded in (23) above.
(28) Set synthesizer/level generator frequency to 1 MHz.
(29) Press FREQUENCY key then [CENTER FREQ] key and enter 1 MHz using
DATA keys.
(30) Press MARKER PEAK SEARCH key.
(31) TI MKR indication will be within r1.8 dB of measuring receiver indication
recorded in (7) above.
(32) Repeat (28) through (31) above for synthesizer/level generator frequency and TI
CENTER FREQ 100 kHz, 10 kHz, 1 kHz, 500 Hz and 200 Hz settings.
b. Adjustments. Refer to paragraph 6 c above.
17. Residual Responses
a. Performance Check
(1) Connect TI CAL OUTPUT to INPUT 50 :.
31