Frequency Response
Table 18. Frequency Response

Calibration Procedure For Spectrum Analyzer Models 8562Ec And 8562Ec-104 Page Navigation
  24    25    26    27    28  29  30    31    32    33  
img
. .
TB 9-6625-2333-24
Figure 5. Frequency response - equipment setup.
(2) Press TI keys and enter values using DATA keys as listed in (a) through
(e) below:
(a) INSTRUMENT STATE PRESET.
(b) FREQUENCY then [CENTER FREQ] to 300 MHz.
(c) SPAN then [SPAN] to 1 MHz.
(d) AMPLITUDE then [LOG dB/DIV] to 2 dB/DIV.
(e) [MORE 1 OF 3] then [COUPLING AC DC] to select DC.
(3) Set synthesized signal generator frequency to 300 MHz and level output to -5 dBm.
(4) Press TI MARKER PEAK SEARCH key.
(5) Set measuring receiver to measure RF power in dBm at 300 MHz.
(6) Adjust synthesized signal generator level output for a TI MKR amplitude
indication between -9.95 and -10.05 dBm.
(7) Record measuring receiver indication.
(8) Set synthesized signal generator frequency to 13.2 GHz.
(9) Set measuring receiver to measure RF power in dBm at synthesized signal
generator frequency.
(10) Press TI FREQUENCY key then [CENTER FREQ] key and enter 13.2 GHz
using DATA keys.
(11) Press MARKER PEAK SEARCH key.
(12) Adjust synthesized signal generator level output for a TI MKR amplitude
indication between -9.95 and -10.05 dBm.
(13) Measuring receiver indication will be within r2.9 dB of indication recorded in
(7) above.
(14) Repeat (8) through (13) above for synthesized signal generator frequency and TI
CENTER FREQ 12.5 GHz to 6.5 GHz settings listed in table 18.