TB 9-6625-2333-24
CENTER FREQ 6.0 GHz to 3.0 GHz settings listed in table 18. Measuring receiver
indication will be within r2.5 dB of indication recorded in (7) above.
CENTER FREQ 2.5 GHz to 20.0 MHz settings listed in table 18. Measuring receiver
indication will be within r1.8 dB of indication recorded in (7) above.
Synthesized signal generator frequency
and test instrument CENTER FREQ
settings
12.5
GHz
12.0
GHz
11.5
GHz
11.0
GHz
10.5
GHz
10.0
GHz
9.5
GHz
9.0
GHz
8.5
GHz
8.0
GHz
7.5
GHz
7.0
GHz
6.5
GHz
6.0
GHz
5.5
GHz
5.0
GHz
4.5
GHz
4.0
GHz
3.5
GHz
3.0
GHz
2.5
GHz
2.0
GHz
1.5
GHz
1.0
GHz
500
MHz
100
MHz
50
MHz
20
MHz
(17) Set synthesized signal generator frequency to 10 MHz.
(18) Set measuring receiver to measure RF power in dBm at 10 MHz.
(19) Press TI FREQUENCY key then [CENTER FREQ] key and enter 10 MHz
using DATA keys.
(20) Press TI SPAN key then [SPAN] key and enter 10 kHz using DATA keys.
(21) Press MARKER PEAK SEARCH key.
30