Table 14. Frequency Readout Accuracy 4 GHz
Figure 4. Frequency span accuracy - equipment setup.

Calibration Procedure For Spectrum Analyzer Models 8562Ec And 8562Ec-104 Page Navigation
  21    22    23    24    25  26  27    28    29    30    31  
img
. . . . .
TB 9-6625-2333-24
MARKER PEAK SEARCH.
(b)
INSTRUMENT STATE AUX CTRL.
(c)
[INTERNAL MIXER].
(d)
[PRESEL AUTO PK] and wait for PEAKING message to disappear.
(e)
SPAN then [SPAN] to 1 MHz.
(f)
(15) Press TI MARKER PEAK SEARCH key. TI MKR frequency indication will be
between 8.999988 and 9.000012 GHz.
(16) Press TI SPAN key then [SPAN] key and enter 10 MHz using DATA keys.
(17) Press TI MARKER PEAK SEARCH key. TI MKR frequency indication will be
between 8.99948 and 9.00052 GHz.
(18) Repeat technique of (16) and (17) above for remaining TI SPAN settings listed
in table 15. TI MKR frequency indications will be within limits specified in table 15.
Table 15. Frequency Readout Accuracy 9 GHz
Synthesized signal
Test instrument
generator
MKR frequency indications
frequency setting
CENTER FREQ
(GHz)
(GHz)
SPAN
setting
Min
Max
settings
(GHz)
9
9
20 MHz
8.99895
9.00105
9
9
50 MHz
8.99745
9.00255
9
9
100 MHz
8.9948
9.0052
9
9
1 GHz
8.950
9.050
(19) Set synthesized signal generator frequency to 1.5 GHz.
(20) Press TI keys and enter values using DATA keys as listed in (a) through
(c) below:
(a) FREQUENCY then [CENTER FREQ] to 1.5 GHz.
(b) SPAN then [SPAN] to 1 MHz.
(c) MARKER FREQ COUNT then [COUNTER RES] to 1 Hz.
(21) Press MARKER PEAK SEARCH key. TI MKR frequency indication will be
between 1.499999997 and 1.500000003 GHz.
(22) Repeat technique of (19) through (21) above for synthesized signal generator and
TI settings listed in table 16. TI MKR frequency indication will be within limits specified
in table 16.