• Home
  • Download PDF
  • Order CD-ROM
  • Order in Print
Table 12. Challenge/Tag Generator - Continued
Table 14. Word B Generator

Calibration Procedure For Transponder Test Set Model An/Upm-155 Page Navigation
  17    18    19    20    21  22  23    24    25    26    27  
img
. . .
TB 9-6625-2337-24
ENTR.
(c)
Use the arrow keys to highlight (WORD A).
(d)
Use the arrow keys to highlight (S1: ON).
(e)
Use the arrow keys to highlight (S2: ON).
(f)
Use the arrow keys to highlight (S3: ON).
(g)
Use the arrow keys to highlight (S4: ON).
(h)
(6) Use the oscilloscope delta time markers to measure the first and second pulses of
WORD A.
(7) Verify that the oscilloscope delta time markers indicate within the limits listed
in first row of table 13.
(8) Repeat technique of (6) and (7) for remaining rows in table 13.
Table 13. Word A Generator
TI
Oscilloscope
time indication
Minimum
Maximum
Test description
( s)
( s)
From pulse
To pulse
1
2
1.93
2.07
WORD A 1 TO 2
1
3
3.93
4.07
WORD A 1 TO 3
1
4
5.93
6.07
WORD A 1 TO 4
1
5
9.93
10.07
WORD A 1 TO 5
1
6
12.93
13.07
WORD A 1 TO 6
1
7
14.93
15.07
WORD A 1 TO 7
1
8
16.93
17.07
WORD A 1 TO 8
1
9
19.93
20.07
WORD A 1 TO 9
1
10
23.93
24.07
WORD A 1 TO 10
1
11
25.93
26.07
WORD A 1 TO 11
1
12
29.93
30.07
WORD A 1 TO 12
1
13
32.93
33.07
WORD A 1 TO 13
1
14
34.93
35.07
WORD A 1 TO 14
1
15
37.93
38.07
WORD A 1 TO 15
1
16
40.93
41.07
WORD A 1 TO 16
1
17
42.93
43.07
WORD A 1 TO 17
1
18
44.93
45.07
WORD A 1 TO 18
1
19
47.93
48.07
WORD A 1 TO 19
1
20
49.93
50.07
WORD A 1 TO 20
1
21
51.93
52.07
WORD A 1 TO 21
1
22
53.93
54.07
WORD A 1 TO 22
1
23
57.93
58.07
WORD A 1 TO 23
1
24
60.93
61.07
WORD A 1 TO 24
1
25
62.93
63.07
WORD A 1 TO 25
1
26
65.93
66.07
WORD A 1 TO 26
1
27
67.93
68.07
WORD A 1 TO 27
1
28
70.93
71.07
WORD A 1 TO 28
(9) Press TI keys as listed in (a) through (c) below:
22


Privacy Statement